VALIDATION and PREVENTIVE MAINTENANCE
Biohazard Safety Cabinet Certification
- Inflow Velocity Test
- Down Flow Velocity Test
- Filter Leak Test
- Site Installation, Alarms Verification Test
- Smoke Pattern Test
Optional Test
- UV Light Radiation Test
- Light Intensity Test
- Noise Level Test
- Vibration Test
- Particle Count Testing
All certification procedures shall be in compliance with the following standards
Laminar Flow Cabinet Certification
- Air Velocity Test
- Particle Count Test
- UV Light Radiation Test (Optional Test)
- Filter Integrity Test
- Light Intensity Test
- Noise Level Test
Certification procedures based on NSF49/Annex F,
EN12469:2000 and ISO14644 (clean room requirements)
Fume Hood Certification
- Face Velocity Test
- Smoke Visualization Test
- Local Visualization Challenge
- Large Volume Visualization Challenge (Optional Test)
- Noise Level Test
- Light Intensity Test
- Activated Carbon Filter Saturation Test
- SF6 Tracer Gas Leak Test
Based on ASHRAE 110-1995 Method of testing performance of laboratory fume hood
Acceptance criteria of face velocity base on ASHRAE 1995 handbook,
SEFA-1-2006 and Industrial ventilation standards
Gaseous Fumigation (Requirements for HEPA filter replacement for BSC)
Supply and Installation of Activated Carbon Filter and HEPA Filter
Temperature Mapping -30 °C to 250 °C
- To
measure the following:
- Temperature Uniformity
- Stability
- Overall Variation
- Indicating Accuracy
- Empty Enclosure Temperature Mapping Test
- Open-door Study or Temperature Recovery Study
- Power Failure or Insulation Thermal Conductivity/ Temperature Change Study
- Full Loaded Enclosure Temperature Mapping Test